High-flow nasal cannula for COVID-19 patients: low risk of bio-aerosol dispersion
Selasa, 19 Mei 2020 08:51:43

Jie Li, James B. Fink, Stephan Ehrmann

European Respiratory Journal 2020 55: 2000892; DOI: 10.1183/13993003.00892-2020

Bio-aerosol dispersion via high-flow nasal cannula shows a similar risk to standard oxygen masks. High-flow nasal prongs with a surgical mask on the patient's face might benefit hypoxaemic COVID-19 patients without added risk for the environment. https://bit.ly/34p7Fyy

To the Editor:

Human-to-human severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) transmission has been established, with >3300 clinicians reported to be infected in China and >1116 clinicians infected in Italy, where 13 882 cases were confirmed by 13 March 2020. Room surfaces in the vicinity of coronavirus disease 2019 (COVID-19) symptomatic patients and clinicians' protective equipment were found to be contaminated. The primary strategy for COVID-19 patients is supportive care, including oxygen therapy for hypoxaemic patients, in which high-flow nasal cannula (HFNC) has been reported to be effective in improving oxygenation. Among patients with acute hypoxaemic respiratory failure, HFNC was proven to avoid intubation compared to conventional oxygen devices. However, there is an important concern that HFNC may increase bio-aerosol dispersion in the environment due to the high gas flow used. The increased dispersion might favour transmission of infectious agents (such as SARS-CoV-2) carried in aerosol droplets generated by the infected patient. This concern is reflected in the limited use of HFNC in the first clinical study reporting 21 patients with COVID-19 in Washington State (USA), where only one patient used HFNC. In contrast, a broad utilisation was observed in the study by Yang et al. from Wuhan, China, where 33 out of 52 intensive care unit (ICU) patients were treated with HFNC.

There appears to be an uncertainty and a trend to avoid HFNC among COVID-19 patients in the western world, thus increasing early intubation rates and potentially associated harms such as sedation and prolonged ICU stay but also intubation procedures per se, which represent a high-risk situation for viral exposure. Early intubation increases the demand for ventilators, contributing to the critical shortage reported worldwide. Avoiding or delaying invasive mechanical ventilation could substantially reduce immediate demand for ventilators. Thus, we aim to discuss the scientific evidence supporting the risk of HFNC-induced bio-aerosol dispersion in the COVID-19 context.

The utilisation of smoke (an aerosol of solid particles <1 µm) simulation via a manikin model by Hui et al.  and Ip et al. provides a direct visualisation of exhaled smoke dispersion. It appears that, when using HFNC, dispersion is greater at 60 L·min−1 than at 10 L·min−1. We summarise the results from reported in vitro studies with different oxygen devices in table 1 . Interestingly, using the same study method and similar breathing patterns, the exhaled smoke dispersion distance from the manikin with HFNC at 60 L·min−1 was similar to the one observed with a simple oxygen mask at 15 L·min−1  and even smaller than with other oxygenation devices, particularly non-rebreathing and Venturi masks. While the dispersion of smoke in this model is instructive, especially between interfaces, the particle size of smoke (<1 µm) only represents a small fraction of the mass of bio-aerosol generated by patients naturally. As the aerosol generated by a patient's cough contains particles from 0.1 to 100 µm, clinical studies are required to truly evaluate aerosol dispersion, particularly the aerosol dynamics during physiological exhalation and cough.

TABLE 1

Summary of exhaled smoke dispersion distances with different oxygen devices

Leung et al. reported a randomised controlled trial comparing the utilisation of HFNC at 60 L·min−1 with an oxygen mask at 8.6±2.2 L·min−1 in 19 ICU patients with bacterial pneumonia on the environmental contamination. The patient's room air was sampled and settle plates were placed at 0.4 m and 1.5 m from patients. No significant difference in bacterial counts was reported in the air sample and settling plates between the two oxygen devices at 1, 2 and 5 days of incubation. These clinical results confirm the in vitro smoke experiments.

In vitro and clinical studies have demonstrated that placing a simple surgical protection mask on patients significantly reduces dispersion distance and levels of virus-infected bio-aerosol 20 cm away from patients while coughing. Such a surgical mask can be worn by a patient oxygenated through a nasal cannula (standard nasal cannula or HFNC) but not when using simple, non-rebreathing or Venturi oxygen masks.

Taken together, compared to oxygen therapy with a mask, the utilisation of HFNC does not increase either dispersion or microbiological contamination into the environment. The patient being able to wear a surgical mask on top of HFNC, in order to reduce the aerosol transmission during coughing or sneezing, represents an additional benefit.

However, given the high efficacy of HFNC to oxygenate the patients, closely monitoring the use of HFNC for COVID-19 patients is crucial to avoid any delay in intubation. Monitoring respiratory rates and pulse oximetry, and clinical examination, are essential.

In conclusion, massive numbers of clinicians have been infected during the COVID-19 outbreak, which has raised concerns around implementing aerosol-generating procedures. Consequently, there appears to be a trend to avoid HFNC. The scientific evidence of generation and dispersion of bio-aerosols via HFNC summarised here shows a similar risk to standard oxygen masks. HFNC prongs with a surgical mask on the patient's face could thus be a reasonable practice that may benefit hypoxaemic COVID-19 patients and avoid intubation.

Clinicians should consider moving away from the dogma restraining the use of HFNC among COVID-19 patients.

Footnotes

  • Author contributions: S. Ehrmann, J.B. Fink and J. Li conceived of the idea. J. Li performed the literature search and drafted the manuscript. All authors reviewed and revised the manuscript and approved the final draft.

  • Conflict of interest: J. Li has nothing to disclose.

  • Conflict of interest: J.B. Fink is the Chief Science Officer of Aerogen Pharma Corp.

  • Conflict of interest: S. Ehrmann reports grants, personal fees and non-financial support from Fisher and Paykel, during the conduct of the study; grants, personal fees and non-financial support from Aerogen Ltd, personal fees and non-financial support from La diffusion technique française, grants from Hamilton medical, outside the submitted work.